A novel method (RIM-Deep) enhances imaging depth and resolution stability of deep-cleared brain tissue in inverted confocal microscopy

  1. Microbiome Medicine Center, Department of Laboratory Medicine, ZhuJiang Hospital, Southern Medical University, Guangzhou, China
  2. Nikon Precision, Shanghai, China
  3. Department of Neurosurgery, ZhuJiang Hospital, Southern Medical University, The National Key Clinical Specialty, Guangzhou, China
  4. State Key Laboratory of Organ Failure Research, Southern Medical University, Guangzhou, China

Peer review process

Not revised: This Reviewed Preprint includes the authors’ original preprint (without revision), an eLife assessment, public reviews, and a provisional response from the authors.

Read more about eLife’s peer review process.

Editors

  • Reviewing Editor
    Paul Barber
    King's College London, London, United Kingdom
  • Senior Editor
    Albert Cardona
    University of Cambridge, Cambridge, United Kingdom

Reviewer #1 (Public review):

Summary:

Liu et al., present an immersion objective adapter design called RIM-Deep, which can be utilized for enhancing axial resolution and reducing spherical aberrations during inverted confocal microscopy of thick cleared tissue.

Strengths:

RI mismatches present a significant challenge to deep tissue imaging, and developing a robust immersion method is valuable in preventing losses in resolution. Liu et al., present data showing that RIM-Deep is suitable for tissue cleared with two different clearing techniques, demonstrating the adaptability and versatility of the approach.

Weaknesses:

Liu et al., claim to have developed a useful technique for deep tissue imaging, but in its current form, the paper does not provide sufficient evidence that their technique performs better than existing ones.

Reviewer #2 (Public review):

Summary:

Liu et al investigated the performance of a novel imaging technique called RIM-Deep to enhance the imaging depth for cleared samples. Usually, the imaging depth using the classical confocal microscopy sample chamber is limited due to optical aberrations, resulting in loss of resolution and image quality. To overcome this limitation and increase depth, they generated a special imaging chamber, that is affixed to the objective and filled with a solution matching the refractive indices to reduce aberrations. Importantly, the study was conducted using a standard confocal microscope, that has not been modified apart from exchanging the standard sample chamber with the RIM-Deep sample holder. Upon analysing the imaging depth, the authors claim that the RIM-Deep method increased the depth from 2 mm to 5 mm. In summary, RIM-Deep has the potential to significantly enhance imaging quality of thick samples on a low budget, making in-depth measurements possible for a wide range of researchers that have access to an inverted confocal microscope.

Strengths:

The authors used different clearing methods to demonstrate the suitability of RIM-Deep for various sample preparation protocols with clearing solutions of different refractive indices. They clearly demonstrate that the RIM-Deep chamber is compatible with all 3 methods. Brain samples are characterized by complex networks of cells and are often hard to visualize. Despite the dense, complex structure of brain tissue, the RIM-Deep method generated high quality images of all 3 samples given. As the authors already stated, increasing imaging depth often goes hand in hand with purchasing expensive new equipment, exchanging several microscopy parts or purchasing a new microscopy set-up. Innovations, such as the RIM-Deep chamber, hence, might pave the way for cost-effective imaging and expand the applicability of an inverted confocal microscope.

Weaknesses:

(1) However, since this study introduces a novel imaging technique, and therefore, aims to revolutionize the way of imaging large samples, additional control experiments would strengthen the data. From the 3 clearing protocol used (CUBIC, MACS and iDISCO), only the brain section from Macaca fascicularis cleared with iDISCO was imaged with the standard chamber and the RIM-Deep method. This comparison indeed shows that the imaging depth thereby increases more than 2-fold, which is a significant enhancement in terms of microscopy. However, it would have been important to evaluate and show the difference of the imaging depth also on the other two samples, since they were cleared with different protocols and, thus, treated with clearing solutions of different refractive indices compared to iDCISCO.

(2) The description of the figures and figure panels should be improved for a better understanding of the experiments performed and the thus resulting images/data.

(3) While the authors used a Nikon AX inverted laser scanning confocal microscope, the study would highly benefit from evaluating the performance of the RIM-Deep method using other inverted confocal microscopes or even wide-field microscopes.

Author response:

Reviewer #1 (Public review):

Summary:

Liu et al., present an immersion objective adapter design called RIM-Deep, which can be utilized for enhancing axial resolution and reducing spherical aberrations during inverted confocal microscopy of thick cleared tissue.

Strengths:

RI mismatches present a significant challenge to deep tissue imaging, and developing a robust immersion method is valuable in preventing losses in resolution. Liu et al., present data showing that RIM-Deep is suitable for tissue cleared with two different clearing techniques, demonstrating the adaptability and versatility of the approach.

Greetings, we greatly appreciate your feedback. In truth, we have utilized three distinct clearing techniques, including iDISCO, CUBIC, and MACS, to substantiate the adaptability and multifunctionality of the RIM-Deep adapter.

Weaknesses:

Liu et al., claim to have developed a useful technique for deep tissue imaging, but in its current form, the paper does not provide sufficient evidence that their technique performs better than existing ones.

We are in complete agreement with your recommendation, and the additional experiments will conduct a thorough comparison of the efficacy between the RIM-deep adapter and the official adapter in the context of fluorescence bead experiments, along with their performance in cubic and MASC tissue clearing techniques.

Reviewer 2 (Public review):

The authors used different clearing methods to demonstrate the suitability of RIM-Deep for various sample preparation protocols with clearing solutions of different refractive indices. They clearly demonstrate that the RIM-Deep chamber is compatible with all three methods. Brain samples are characterized by complex networks of cells and are often hard to visualize. Despite the dense, complex structure of brain tissue, the RIM-Deep method generated high-quality images of all three samples. As the authors stated, increasing imaging depth often goes hand in hand with purchasing expensive new equipment, exchanging several microscopy parts, or purchasing a new microscopy setup. Innovations like the RIM-Deep chamber might pave the way for cost-effective imaging and expand the applicability of inverted confocal microscopy.

Weeknesses:

(1) However, since this study introduces a novel imaging technique aiming to revolutionize imaging of large samples, additional control experiments would strengthen the data. From the three clearing protocols used (CUBIC, MACS, and iDISCO), only the brain section from Macaca fascicularis cleared with iDISCO was imaged with the standard chamber and the RIM-Deep method. This comparison indeed shows a more than 2-fold increase in imaging depth, a significant enhancement in microscopy. However, it would have been important to evaluate and show the imaging depth differences in the other two samples, as they were cleared with different protocols and treated with clearing solutions of different refractive indices compared to iDISCO.

Thank you for your suggestion. We will investigate the imaging performance of brain tissue using the other two clearing protocols with both the official adapter and the RIM-deep method.

(2) The description of the figures and figure panels should be improved for a better understanding of the experiments performed and the resulting images/data.

Thank you for your suggestion. We will revise the figure legends in detail.

(3) While the authors used a Nikon AX inverted laser scanning confocal microscope, the study would benefit from evaluating the performance of the RIM-Deep method using other inverted confocal microscopes or even wide-field microscopes.

Thank you for your suggestion. We also recognize that evaluating the performance of the RIM-Deep method on other inverted confocal microscopes will help further validate its applicability and robustness. We will supplement these experiments to expand the scope and reliability of RIM-Deep.